The Apollo system is designed to be safe for individuals with pacemakers and implantable cardioverter defibrillators (ICDs). Its electromagnetic fields (EMFs) are kept below the limits set by the ISO 14117:2019 standard, which is specifically designed to prevent interference with these devices. 


ELECTROMAGNETIC FIELD LIMITS 

Frequency Range: 0 Hz ≤ ƒ < 385 MHz 

Induced Lead Current: 

  • 16.6 Hz ≤ ƒ ≤ 1 kHz: Maximum current induced in the lead is 50 µA RMS. 
  • 1 kHz ≤ ƒ ≤ 20 kHz: Maximum current is 50 µA × (ƒ / 1 kHz). 

Voltage Induced in Leads: 

  • 16.6 Hz ≤ ƒ ≤ 20 kHz: Maximum voltage is 1 V peak-to-peak. 
  • 20 kHz ≤ ƒ ≤ 140 kHz: Maximum voltage is 1 V × (ƒ / 20 kHz). 
  • 140 kHz ≤ ƒ ≤ 10 MHz: Maximum voltage is 7 V × (ƒ / 140 kHz)^0.1624. 

Frequency Range: 385 MHz ≤ ƒ ≤ 3 000 MHz 

  • Proximity Fields at 15 cm Separation: 
    • Net dipole power: 120 mW RMS. 

Optional Characterization Testing: 

  • 385 MHz ≤ ƒ < 1 000 MHz: Net dipole power of 8 W RMS. 
  • 1 000 MHz ≤ ƒ ≤ 3 000 MHz: Net dipole power of 2 W RMS. 

Frequency Range: ƒ ≥ 3 000 MHz 

  • Testing above 3 GHz is not required due to reduced sensitivity and shielding effects. 

STATIC MAGNETIC FIELDS 

Flux Density up to 1 mT: 

  • The device remains uninfluenced and continue functioning normally. 

Flux Density up to 50 mT: 

  • This device recovers its original functionality after exposure without adjustment. 

info@apollosecurity.us

AC MAGNETIC FIELDS (1 KHZ TO 140 KHZ) 

Magnetic field strength: 

  • 1 kHz ≤ ƒ ≤ 100 kHz: 150 A/m RMS. 
  • 100 kHz ≤ ƒ ≤ 140 kHz: 150 A/m × (100 kHz / ƒ). 

CONTINUOUS WAVE (CW) SOURCES 

Frequency Range: 16.6 Hz to 167 kHz: 

  • Maximum test signal amplitude: 
    • Unipolar/Common Mode: 1 V peak-to-peak. 
    • Bipolar Differential Mode: 0.1 V peak-to-peak. 

RESULTS 

Measure EMF Emissions Results: 

  • Apollo’s electromagnetic field strength stays below the limits specified above.
  • Frequencies Below 1 kHz: PASS (These frequencies are particularly sensitive for pacemakers and ICDs ) 

Email- support@athena-security.com,

Call-833-928-4362  27/7 Support

Athena Security Help Desk Ticket (website) 

info@apollosecurity.us 

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article